Model 9000 Total Hydrocarbon Analyzer

The Model 9000 is based on a FlowGuard controlled flame ionization detector (FID) and has an incredible range from less than 10 ppb on the high sensitivity version to 100% on the percent level version. The analyzer has a generous complement of analog, digital, and logic output capabilities. It can be configured with internal components for a single or multipoint analysis of non-condensing gas samples.

  • Automatic calibration
  • FlowGuard control of fuel, air, and sample
  • Graphical display and easy to use menu system
  • Programmable relays
  • Automatic FID ignition

Data Sheet - English
Request A Quote


Model 9000 Analyzer - Product Specification

Sampling: Internal single or multipoint modules, with or without sample pump(s) for prefiltered (< or = .1 microns), non-condensing samples
Detector: Flame ionization detector (FID)
Calibration: Programmable automatic, or manual (with internal selection valves)
MDQ: Minimum detectable quantity: Low Range (0.01 ppm ), Med Range (0.10 ppm), High Range (10.0 ppm).
Range: Low Range (0-2,000 ppm Methane), Med Range (0-20,000 ppm Methane), High Range (0-100% Methane)
Repeatability: ± 1% Full-scale response
Drift: ± 1% Over 24 hours
Response Time: < 5 Seconds to 90% of final reading
Display: 3.4" x 4.5" graphical display
Output: Analog - 1 programmable 0-20mA or 4-20mA isolated output.

Digital - RS-232, Ethernet
Relays: 5 programmable form A relays rated to 3A @ 230V.
Alarms: Multilevel concentration and fault alarms that result in an audible and visually displayed alarm, can be mapped to relays to control external equipment.
Physical: Dimensions: 19.00" W x 14.25" D x 5.25" H (48.26 cm W x 36.195 cm D x 13.335 cm H).
Nominal weight: < 20 lb (9.072 kg).
Configuration: Bench-top or rack-mount (19" panel)
Gas Specifications: Support - Hydrocarbon content:<1 ppm required. Air @ approx. 200 cc/min, hydrogen @ approx. 40 cc/min. (Options: H2/N2 or H2/He @ 100 cc/min.)

Connections - 1/4" O.D. Tube fitting connectors (1/8", 4 mm, and other options)
Power: 90-230 VAC,, 50/60 Hz, 3A
Operating Conditions: Temperature: 32-104 °F (0-40 °C). Humidity: 0-95%, non-condensing.

Model 9000 Analyzer - Options & Accessories

Samplers: Internal multipoint module, with or without sample pump available in 4-point
Enclosures: General purpose, Wallmount, X-purged or Z-purged
Expansion Boards: Analog - Provides up to 10 additional programmable 4-20 mA outputs, with sample read & hold

Relay - Provides up to 10 additional programmable relays
Catalysts: Methanizer for low-level CO/CO2 analysis
Oxidizer for methane only analysis
Gas Generators: Zero air or hydrogen
Calibration Gas: Zero and span gases for a variety of applications