Model 8800 PID Total Volatile Organic Vapor Analyzer

The Model 8800 PID has a photoionization detector (PID) and an incredible dynamic range from 10 ppb to 1%. It is designed to analyze hundreds of volatile organic compounds (VOCs) and various other gases. The analyzer is relatively humidity insensitive and can be configured with internal components for a single or multipoint analysis of noncondensing gas samples.

  • Automatic calibration
  • Programmable virtual analog ranges
  • Programmable relays
  • RS-485, RS-232 output


Data Sheet - English
Data Sheet - Chinese
Request A Quote


Model 8800 PID Analyzer - Product Specification

Sampling: Internal single or multipoint modules, with or without sample pump(s), for prefiltered (< or = .1 microns), non-condensing samples
Calibration: Programmable automatic, or manual (with internal selection valves)
Detector: Photoionization detector (PID)
Lamp: Energies: 10.6 eV (life span >6000 hrs), 11.7 eV (life span approximately 140 hrs).
MDQ: Minimum detectable quantity: <0.1 ppm (as isobutylene), <0.1 ppm (as benzene).
Quenching: Signal quenching due to moisture:< 30% at 95% R.H. and 23° Celsius.
Range: Analog - Virtual range with software selectable endpoints provides full-scale ranges from 1.0 ppm - 1% (as isobutylene)

Digital - Display auto-ranges from 0.01 ppm to 1% (as isobutylene)
Linearity: Linear range: 0 - 10,000 ppm (isobutylene). Accurate to ± 1 ppm or ± 15% of reading, whichever is greater.
Drift: Sample dependent. Zero:< 0.1 ppm (as isobutylene) over 24 hours. Span: 100 ppm isobutylene,< 3 % over 24 hours.
Response Time: Isobutylene:< 6 Seconds to 90% of final reading
Output: Analog - 1 (standard) to 15 programmable 0-20 mA or 4-20 mA loop power supplied, isolated outputs or  optional 0-1 V, 0-5 V, or 0-10 V isolated outputs. Selectable for concentration, temperature, flow (fuel, air, or sample).

Digital - Standard: RS-485 output (RS-232 option).
Relays: 5 (standard) to 15 programmable (latched/not, NO/NC) contact closures (1A@30V max).  Selectable for alarm thresholds or events (calibration, fault, or sample location).
Alarms: Multilevel concentration, average concentration, and fault

Audible - Horn: Sounducer, generating 85 dB @10 cm. Selectively en-/disabled for keypad input, fault, and alarms.
Physical: Dimensions: 19.00" W x 16.00" D x 8.75" H (48.26 cm W x 40.64 cm D x 22.23 H). Nominal weight: 30 lb (13.64 kg).
Configuration: Bench-top or rack-mount (19" panel)
Display: Digital vacuum fluorescent, 20 characters x 2 lines
Power: 90-120 VAC or optional 210-230 VAC,  50-60 Hz
Operating Conditions: Temperature: 32-104 °F (0-40 °C). Humidity: 0-95%, non-condensing.
Gas Specifications: Span - Isobutylene, or as required by application

Connections - 1/4" O.D. Tube fitting connectors (1/8", 4 mm, and other options)

Model 8800 PID Analyzer - Options & Accessories

Samplers: Internal multipoint module, with or without internal sample pump(s), available in 4-point or 8-point configurations
Enclosures: General purpose, X-purged or Z-purged
Expansion Boards: Analog - Provides 4 or 10 additional programmable 4-20 mA outputs, with sample read & hold

Relay - Provides up to 10 additional programmable relays
Calibration Gas: Zero and span gases for a variety of applications

Model 8800 PID Analyzer - Technical Drawing