Model 8800 H Heated Total Hydrocarbon Analyzer

The Model 8800 H is based on a flame ionization detector (FID) and maintains the temperature of heated samples up to 200 °C. It has a dynamic range from less than 0.1 ppm to 100%. The analyzer is equipped with a generous complement of analog, digital, and logic output capabilities with room to expand.

  • Automatic calibration
  • Electronic flow control of fuel, air & sample
  • Sample pump starts automatically after warm-up
  • Programmable virtual analog ranges
  • Programmable relays
  • Automatic FID ignition


Data Sheet - English
Data Sheet - Chinese
Request A Quote


Model 8800 H Analyzer - Product Specification

Sampling: Heated, single point sample train, for prefiltered (< or = .1 microns) samples
Calibration: Programmable automatic, or manual with optional internal selection valves
Detector: Flame ionization detector (FID)
MDQ: Minimum detectable quantity: 0.1 ppm (100 ppb) as propane
Range: Analog - Virtual range with software selectable endpoints provides full-scale ranges.  Low range unit with -50V collector, provides .1 ppm - 10% (as propane).  High range unit with -300V collector, provides 1 ppm - 50% (as methane)

Digital - Display auto-ranges from 0.1 ppm to over 50%
Repeatability: ± 1% Full-scale response
Drift: ± 1% Over 24 hours
Response Time: < 5 Seconds to 90% of final reading
Output: Analog - 1 (standard) to 15 analog 0-20 mA or 4-20 mA loop power supplied, isolated outputs or optional 0-1V, 0-5V, or 0-10V isolated outputs.  Selectable for concentration, temperature, flow (fuel, air, or sample).

Digital - Standard: RS-485 output (RS-232 option).
Relays: 5 (standard) to 15 programmable (latched/not, NO/NC contact closures (1A@30V max).  Selectable for alarm thresholds or events (calibration, fault, or sample location).
Alarms: Multilevel concentration, average concentration, and fault

Audible - Horn: Sounducer, generating 85 dB @10 cm. Selectively en-/disabled for keypad input, fault, and alarms.
Physical: Dimensions: 19.00" W x 16.00" D x 8.75" H (48.26 cm W x 40.64 cm D x 22.23 H). Nominal weight: 45 lb (20.5 kg).
Configuration: Bench-top or rack-mount (19" panel)
Display: Digital vacuum fluorescent, 20 characters x 2 lines
Power: 90-120 VAC or optional 210-230 VAC, 50-60 Hz
Operating Conditions: Temperature: 32-104 °F (0-40 °C). Humidity: 0-95%, non-condensing.
Gas Specifications: Support - Hydrocarbon content:<1 ppm required. Air @ approx. 200 cc/min, hydrogen @ approx. 40 cc/min. (Options: H2/N2 or H2/He @ 100 cc/min.)

Connections - 1/4" O.D. Tube fitting connectors (1/8", 4 mm, and other options)

Model 8800 H Analyzer - Options & Accessories

Samplers Internal multipoint modules, with or without sample pump(s) available in 4-point or 8-point configurations.
Enclosures: General purpose, wallmount, X-purged or Z-purged
Expansion Boards: Analog - Provides 4 or 10 additional programmable 4-20 mA outputs, with sample read & hold

Relay - Provides up to 10 additional programmable relays
Gas Generators: Zero air or hydrogen
Calibration Gas: Zero and span gases for a variety of applications

Model 8800 H Analyzer - Technical Drawing